Lab Instruments

MT-1001

MT-1001

Multi Tester can measure most of the important factors for powder evaluation, particle design, and mill system and powder transportation system design. For the analysis and the characteristic values of measurement, Kawakita equation and Carr's index are used. MT-1001 is also applicable to bulk density measurement complying with JIS standards.

Feature

  • For the bulk density measurement, the measurement mehods complying with the following JIS standards can be applied.
    JIS-Z-2504:Metal powder, JIS-K-5101:Pigment, JIS-K-6721:PVC, JIS-K-6891:Resin molding compounds, JIS-R-6126:Artificial abrasives, JIS-K-7365:Plastics(ISO-60) For the tap density measurement complied with United States Pharmacopeia(USP-24)
  • USP-24 is equivalent to the Japanese pharmacopoeia. Tap density measurement method based on USP-24 is required when exported to the USA.
    MT-1001 complies with USP-24 by attaching optional parts.
  • Adhesive force and flowdability can be evaluated from Kawakita's equation.
  • Single sample slot prevents samples from scattering and makes cleaning with vacuumm cleaner easy.
  • Measurement methods and conditions can be easily changed.
  • Reproductivity and accuracy are very high and variation from measurer to measurer in the measurement results is little.
  • Measurement status can be seen through the front transparent acrylic door with the light inside.
  • Sieve vibration strength is adjustable.

Specification

Name & Model

Name Multiple Powder Characteristics Analyzer(Multi Tester)
Model MT-1001
Main Body Power Voltage AC100V±10%(50/60Hz)・3A(Max15A)
Size 600mm(W)×360mm(L)×610mm(H)
Weight About 40kg
Tapping Unit Number of Tapping 2/sec(50Hz)、2.4/sec(60Hz)
Set Number of Tapping Max 9,999
Total Number of Tapping to be Displayed Max 999,999
Tapping Stroke 50mm,40mm,30mm,20mm,
18mm, or 10mm in combination of spacers
Tapping Unit Back-and-forth movement by variable speed indusction motor and roller cam
Reaction Prevent Damper Magnetic damper operation by timing cam syncronization
Cell Holding Method 90 degree turn holding method
Repose Angle Unit Table Diameter 80φ
Table Height 126mm
Load Weight 160g
Load Drop Distance 88mm
Angle Measurement Unit Range 0 - 90 degree
Minimum Scale Value 1 degree